INR online

Results Notification sheet

Dr/NP notification letter - for new patient

Patient Consent Form - for new patient

Add Dr to INR online

Ordering

Name of machine CoaguChek Pro II

Test strips - Item# 172389 - UPC# 4015630933488

Lancing Device - Item# 797209 - UPC 064562213261

Super Sani Cloths - For cleaning INR machine - Easiest to call in 1-800-565-0765.

Control Solution - Item# 172391 - UPC 4015630933235 Refrigerated

Instructions on Control Testing

-Find a box in fridge that says Coagucheck PT Controls - Take out a vial, pipette, and code chip

-Take Coagucheck machine and put in new code chip

-Login —> Control Test —-> Control Test —> Insert regular test strip —> select the new number for the code chip—> See vial for level/control

-Cut pipette at the end and add water to the vial. Close bottle and swirl until dissolved.

-Add solution same way you would a patient sample

-Replace the code chip and put testing chip back in box

Cleaning Instructions

Put on gloves and use the Super Sani-cloths (see above). Use gloves and ring out moisture if cloths are too wet. Clean inside the test strip guide as well. Operators manual recommends at least monthly cleaning.